immerse a 100mm length of tubing in the chemical in question for 72 hours at room temperature. If after
this time period the tubing has increased or decreased in length by less than 3%, the combination is
compatible. Length changes of 3% to 8% are worth trying, but will not produce the most accurate results.
Changes in length over 8% are considered to be unsatisfactory. It should also be noted that a new tubing
assembly may cause slight filling variations for the first one to two hours of operation.
7.3 Switching the Unit ON
Switch the pump ON via the power switch on the rear panel. The main screen displays the Wheaton
Industries logo and the firmware version level.
The pump will automatically transition to the
HOME
screen. Additional screens feature a similar layout:
7.4 Keypad Functions and Icons
Icons on the right of each screen show the user which keys will be active during function and setup
navigation. A brief explanation of active keys is listed:
Help
– Pushing the help button will show icon explanations in several languages
Run/Stop
– Starts and stops the pump
Prime
– Push and hold the button to run the pump at a fixed speed to purge air
from the tubing
Arrow Keys
– Used to navigate menu items, and adjust parameters and values
OK
– Confirms an entry of a parameter or value
Active Keys Available
Screen Name
Function Icons (Highlighted)