16.1 Maintenance
Routine user maintenance should include a regular inspection of the tubing assembly for excessive wear
and leaks. Many service problems can often be cured by simply replacing a worn tubing assembly. To
insure high accuracy, the pump rotor and tubing clamping mechanics of the pumphead should be kept as
clean as possible at all times.
The unit cabinetry has been designed to an IP31 rating. The plastic top is mineral filled polypropylene and
the base is painted cast aluminum. The pumphead and outer cabinet may be wiped down with warm
water and any mild household cleaning solution. Never use abrasive cleaners or strong acids as cleaning
solutions. Never place liquid containers on top of the OmniSpense
®
ELITE as a spill could result in liquid
entering the dispenser. If liquid should accidentally contact the front panel, the unit should be
disconnected from the AC power source immediately and carefully dried with towels and warm air.
Although the OmniSpense
®
ELITE is designed for easy service, it is highly recommended that the
instrument be returned to Wheaton Industries should any major repairs be needed.
Pump rollers can be occasionally lubricated with Teflon lubricating oil.
16.2 Troubleshooting – For Qualified Maintenance Personnel Only!
DANGER! NEVER ATTEMPT TO PERFORM REPAIRS IF THIS INSTRUMENT IS
PLUGGED IN! TO AVOID SERIOUS ELECTRIC SHOCK OR ELECTROCUTION, THIS
INSTRUMENT MUST BE DISCONNECTED FROM THE SOURCE OF AC POWER
BEFORE REMOVAL OF ANY PROTECTIVE COVERS.
Unit will not operate:
Cause:
Fuse blown.
Remedy:
Replace fuse with proper size and type. (see ‘Fuse Replacement’)
Cause:
Supply voltage low or at zero.
Remedy:
Check house receptacle with a voltmeter.
Cause:
On/Off power switch in "off" position.
Remedy
: Switch power switch to "on" position.
Motor runs but pump head(s) will not turn;
Cause:
Pump head not properly engaged with drive shaft.
Remedy:
Realign pump head with drive shaft. (see ‘Initial Startup and Operation)
Switch on but motor fails to rotate;
Cause:
Pump Speed set too low.
Remedy:
Increase pump motor speed as required.
Cause:
Motor control board defective.
Remedy:
Replace defective board.
Cause:
START Key not pressed.
Remedy:
See ‘Unit Features and Component Identification’ for keypad and display operation.