Press
OK
to navigate to the
CALIBRATE
screen.
10.1 Setting Calibration Speed
Use the
LEFT
and
RIGHT
buttons to select the calibration speed:
Slow calibration speed
Fast calibration speed
10.2 Priming the Tubing
Use the
UP
and
DOWN
buttons to select the
PRIME
icon. The PRIME button on the keypad now
becomes active. Push and HOLD the prime button on the keypad to run the pump motor and purge any
air from the tubing. RELEASE the prime button to stop the motor.
10.3 Calibration Dispense
Use the
UP
and
DOWN
button to navigate the
Calibration Dispense
icon:
Prepare to measure a dispense by using a graduated cylinder, electronic scale, or other device that can
accurately measure volume. Press the
START / STOP
button to begin a
calibration dispense.
For
best results, the pump rotor should make at least ten (10) revolutions.
Calibrate Screen
Pump Running /
Stopped Icon
Active Keys