7.5 Footswitch Operation and Connector Functions
Plug the optional footswitch into the Interface Connector found at the rear of the unit. If the unit is at idle,
pressing the footswitch once, will initiate a pump dispensing sequence, as defined by the entries made by
the user via the main front panel keypad. If the unit is running, pressing the footswitch once will stop the
pump.
In addition to connections for the above footswitch, the
interface Connector contains feedback pinouts for a Pump
Running / Pump Stopped indicator.
Pins 1 (footswitch +) and 2 (footswitch GND) are +5V TTL
compatible inputs with pull up resistors, and can be driven by
an isolated open collector device or dry contact relay, to start
and stop the unit.
Pins 3 and 5 are an isolated transistor output capable of driving
a TTL load with pull up resistor. When the pump is
running
,
the transistor is
ON
, and when the pump is
stopped
, the
transistor is
OFF
.
See Figure F for connector pin out details.
7.6 Flexible Tubing Holder
An included flexible tubing holder mounts to either side of the
unit, and can be flexed to the desired position, for hands-free
operation of unit.
8.0 Unit Setup
The setup menu contains a collection of lesser used functions of the unit. From the
HOME
screen, use
the
UP
and
DOWN
keys to navigate to the
SETUP
icon, and then press
OK
.