11.2 Batch Count
Use the
UP
and
DOWN
keys to navigate to the batch count:
Use the
UP
and
DOWN
and the
LEFT
and
RIGHT
keys to enter the batch count. Press
OK
when done.
11.3 Volume per Dispense
Use the
UP
and
DOWN
keys to navigate to the Volume per Dispense icon:
Use the
UP
and
DOWN
and the
LEFT
and
RIGHT
keys to enter volume per dispense. Press
OK
when
done.
11.4 Delay Between Dispenses
Use the
UP
and
DOWN
keys to navigate to the Delay icon:
Use the
UP
and
DOWN
and the
LEFT
and
RIGHT
keys to enter the delay between dispenses. Press
OK
when done.
11.5 Drawback Function
Use the
UP
and
DOWN
keys to navigate to the Drawback icon:
The drawback feature will make the pump
over-dispense
the volume set for the drawback volume, then
reverse
and draw back the same volume set for the drawback. This feature is typically used to fill
intravenous bags, or other unvented vessels. See the example below for a dispense set at 100mL and a
drawback of 5mL: