2. Electrical measurement – Replacement leakage current measurement
(in accordance with VDE 0751 part 1)
•
Put the line cord of the health care bed into the checking socket of the measuring instrument.
•
Connect the probe head of the measuring instrument to a not isolated, conductive part of the
lying surface frame.
•
During the measuring period the manual control unit has to be operated ( refer to 6.3).
•
Start measuring procedure at the measuring instrument.
•
insulating resistance measurement: Initiate insulation test voltage and measure resistance
(Result must be > 10 M Ohm)
•
upper limiting value of appliance’
s conduction current (appliance above 200 VA, min. protection
class II, Type B
(according to VDE 0751-1) :
0,500 m A