Operating instructions Medial
__________________________
32
Central drive unit
:
7 kg (9, 5 kg with rechargeable battery for Trendelenburg
positioning)
Side rails
: 4 kg each
Sit up device : 7,8 Kg (safe workload 75 Kg)
8.3 Castor types: TENTE
Machine castor
:
Diameter 100 mm
Safe work load of the health care bed amounts to : 170 kg
Maximum patient weight :
135 kg
9.0 Electrical data:
9.1 Central drive unit
Type
: Quadromat 7 Care light + shorting-plug (Nr. 4056) from
Firma Dewert
Input voltage
: 230 V AC; 50 Hz
Power consumption
: max. 0, 8 A
Output voltage
: 24 V
Power-on time
: AB: 2 min. On / 18 min Off
Protection class
: IP X4 fulfilled and IP 44 fulfilled
Classification
: Protection class II (protective insulation),
Not for operation in atmospheres which bear a danger of explosion
Line cord
: Spiral wound special wires EPR (
E
thylene
P
ropylene
R
ubber)
in accordance to EN 60601-2-38
Power
: 2x 3500 N
9.2 Integrated front module engines:
Type : Megamat MBZ – additional drive units 2 pieces (No 37713) from
Firma Dewert
Input voltage
: 24 V
Power-on time
: AB: 2 min. On / 18 min Off
Protection class
: IP X4 fulfilled and IP 44 fulfilled
Power : 1500 N
9.3 Noise level: < 65 dB (A)
Diagram of connections, spare part lists as well as further technical details can be obtained
through our service hotline ( refer to page 3).
10.0 Maintenance- checklist:
This inspection consists of a visual control for damages and a function control