3 Description
Battery status approx. 10 % - 40 %
Battery status < 10 %
In the display:
• The last remaining segment in the battery status
symbol is red.
•
Battery weak
appears in the display.
• In AED mode, the device outputs:
Battery weak
.
Battery is deeply discharged. Charge battery in the
device for 24 h. After 24 hours:
• Green LED is on: Battery fully charged and ready
for use
• Red LED or no LED on: Battery defective. Replace
battery.
Battery is empty
Battery empty
appears in the display and in AED
mode, the device outputs:
Battery empty
.
The device can be used for about another 15 minutes.
Battery is defective. Replace battery.
• Battery is defective.
or
• No battery.
or
• Battery not at suitable temperature.
Green arrow: Battery charging
Status indicator on
the battery
Status indicator on
the device display
Meaning