11 Alarms and faults
Service data are not uploaded to
WEINMANN Emergency
WiFi is deactivated
Activate network
Airplane mode is activated
Deactivate airplane mode
Unable to connect network
Bring device within range of the
network. If necessary, activate a
mobile network.
Configure default network
.
Check whether network is
connected to the Internet.
Check whether port 443 is
enabled.
Request connection data for the
network from the system
administrator.
Device not registered with
WEINMANN Connect
Register the device with
WEINMANN Connect
E-mail with 12-lead ECG does not
reach the recipient, or does so with
a severe delay
E-mail with 12-lead ECG lands in
spam/junk folder of the recipient's
mailbox
Add sender’s address to list of safe
senders. If necessary: Contact
system administrator.
E-mail with 12-lead ECG blocked/
delayed by recipient's server
E-mail does not contain a 12-lead
ECG in the attachment
E-mail with 12-lead ECG blocked/
delayed by recipient's server
Add sender’s address to list of safe
senders. If necessary: Contact
system administrator.
Fault
Cause