WM
6840
1
0
4
/2
021
MEDUCORE Standard
2
EN
319
13 Operator menu
Infant alarm limit
settings
Heart rate
q
35/min - 250/min,
in increments of 5
Here you can set the upper
( ) and lower ( ) limit
values for the patient group
as of which the device is to
emit an alarm.
According to the values set,
the setting ranges of the
upper and lower limit values
depend on each other as
follows:
• The setting range for the
upper alarm limit ends at
least one setting value
above the set value for
the lower alarm limit.
• The setting range for the
lower alarm limit ends at
least one setting value
below the set value for
the upper alarm limit.
200/min
Heart rate
r
30/min - 245/min,
in increments of 5
100/min
Pulse rate
q
35/min - 250/min,
in increments of 5
200/min
Pulse rate
r
30/min - 245/min,
in increments of 5
100/min
SpO
2
saturation
q
66 % - 100 %
95 %
SpO
2
saturation
r
65 % - 99 %
85 %
NIBP systolic
q
45 mmHg -
130 mmHg,
in increments of 5
100 mmHg
NIBP systolic
r
40 mmHg -
125 mmHg,
in increments of 5
50 mmHg
NIBP diastolic
q
25 mmHg -
100 mmHg,
in increments of 5
70 mmHg
NIBP diastolic
r
20 mmHg -
95 mmHg,
in increments of 5
30 mmHg
Pause audio
-
1 min
2 min
5 min
10 min
∞
(infinite)
Here you can set the length of
time for which audio alarm
output is paused. If you select
∞
(infinite), audio alarm
output is paused
permanently (audio alarm
output is muted).
2 min
Audible alarm
emitted in
Manual mode
(only with
Manual
defibrillation
option)
-
Activated
Deactivated
Here you can set whether
audio alarm output is active
or inactive when the user
calls up manual mode.
Deactivated
Parameter
Possible
values
Description
Factory