11 Alarms and faults
NIBP signal disrupted
(Alarm deactivated after 10 s)
Unable to detect pulse wave signal
for NIBP measurement reliably
• Make sure that the patient lies
still during the measurement.
• Rule out the risk of patient
vibration.
• Use a suitable size of NIBP
cuff.
• Ensure that the NIBP cuff is
properly attached.
• Ensure that there is no
clothing between the NIBP
cuff and the patient.
NIBP systolic
q
(Alarm deactivated after 10 s)
Measured systolic blood pressure
is above the set alarm limit
If medically indicated, carry out
suitable treatment.
Adapt alarm limits
NIBP systolic
r
(Alarm deactivated after 10 s)
Measured systolic blood pressure
is below the set alarm limit
If medically indicated, carry out
suitable treatment.
Adapt alarm limits
NIBP tube blocked
(Alarm deactivated after 10 s)
NIBP connecting tube kinked or
blocked
Check NIBP connecting tube and
replace if necessary.
NIBP tube leaking
(Alarm deactivated after 10 s)
NIBP connecting tube or NIBP cuff
are leaking or incorrectly
connected. Sufficient pressure
cannot be built up.
Check NIBP connecting tube and
NIBP cuff and replace if necessary.
No printer connected
(Alarm deactivated after 10 s)
Connection between device and
printer interrupted
Wait 10 seconds.
Switch printer on again.
Position the printer within range of
the device.
Plug in ECG connector
ECG connector of the ECG cable
not connected/not connected
correctly to the device
Connect ECG connector of the
ECG cable to the device correctly
Plug in Pad connector
(Monitor mode)
Defibrillation electrodes not
connected/not connected correctly
to the master cable
Connect defibrillation electrodes
to the master cable correctly
Paddles not connected/not
connected correctly to the master
cable
Connect paddles to the master
cable correctly
Master cable not connected/not
connected correctly to the device
Connect the master cable to the
device correctly
Alarm
Cause
Remedy