384
EN MEDUCORE
Standard
2
WM
6840
1
0
4
/2
021
18 Technical data
Electromagnetic emission
Electromagnetic immunity
Electromagnetic emission
MEDUCORE Standard
2
is designed for operation in the electromagnetic environment described below. The
operator or the user must ensure that he or she operates MEDUCORE Standard
2
in an environment of this
kind.
Emission
measurements
Compliance
Electromagnetic environment guidelines
HF emissions according
to CISPR 11
Group 1
MEDUCORE Standard
2
uses high-frequency energy
exclusively for internal functions. As the high-
frequency emission is very low, interference with
electronic devices in the vicinity from
MEDUCORE Standard
2
is unlikely.
Class B
MEDUCORE Standard
2
is suitable for use in the
following areas:
• In all facilities, including those in residential
settings which are connected directly to a
public power supply network
• In automobiles, airplanes and ships
Electromagnetic immunity
MEDUCORE Standard
2
is designed for operation in the electromagnetic environment described below. The
operator or the user must ensure that he or she operates MEDUCORE Standard
2
in an environment of this
kind.
Interference immunity
tests
IEC 60601 test level
Compliance level
Electrostatic discharge (ESD) acc.
to IEC 61000-4-2
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV air
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV air
Radiated HF interference acc. to
IEC 61000-4-3
20 V/m
80 MHz to 2.7 GHz
20 V/m
Rapid transients/bursts acc. to
IEC 61000-4-4
± 2 kV
100 kHz spike frequency
± 2 kV
100 kHz spike frequency
Surges acc. to IEC 61000-4-5
± 0.5 kV, ± 1 kV conductor/
conductor
± 0.5 kV, ± 1 kV, ± 2 kV
conductor/ground
± 0.5 kV, ± 1 kV conductor/
conductor
± 0.5 kV, ± 1 kV, ± 2 kV
conductor/ground