11 Alarms and faults
Pulse rate
q
Measured pulse rate is above the
set alarm limit
If medically indicated, carry out
suitable treatment.
Adapt alarm limits
.
Pulse rate
r
Measured pulse rate is below the
set alarm limit
If medically indicated, carry out
suitable treatment.
Adapt alarm limits
.
Server not accessible
(Alarm deactivated after 60 s)
Server not accessible as there is no
Internet connection
Check using an alternative
network whether the configured
network is the cause.
Settings lost
Settings had to be reset to factory
settings.
Re-enter presets in the operator
menu. If this recurs: Have the
device repaired.
SpO
2
q
Measured oxygen saturation is
above the set alarm limit and SpO
2
signal quality is
≥
40 %
If medically indicated, carry out
suitable treatment.
Adapt alarm limits
.
SpO
2
r
Measured oxygen saturation is
below the set alarm limit and SpO
2
signal quality is
≥
40 %
If medically indicated, carry out
suitable treatment.
Adapt alarm limits
.
SpO
2
module defective
Internal SpO
2
module defective
Have the device repaired.
Unknown defibrillation
electrodes
Unknown defibrillation electrodes
connected to the master cable
Only use defibrillation electrodes
from WEINMANN Emergency.
Unknown paddles connected to
the master cable
Only use paddles from
WEINMANN Emergency.
Upload canceled
(Alarm deactivated after 60 s)
Data transmission canceled by user
• Check date/time
.
.
Data transmission interrupted
during transmission
• Check date/time
.
• Check connection settings
.
.
Upload failed
Connection to network interrupted
. If necessary: Select a
different network.
Alarm
Cause
Remedy