If hazardous fluids are to be pumped, safety procedures specific to the
particular fluid and application must be put in place to protect against injury
to persons.
This product does not comply with the ATEX directive and must not be used in
explosive atmospheres.
Ensure the chemicals to be pumped are compatible with the pumphead,
lubricant (where applicable), tubing, pipework and fittings to be used with
the pump. Please refer to the chemical compatibility guide which can be
found at:
. If you need to use the pump with any
other chemical please contact Watson-Marlow to confirm compatibility.
If the Auto restart feature is enabled it may cause the pump to start as soon
as power is turned on.
Auto-restart only affects operation of manual mode, network mode and
memodose mode.
If Auto restart is enabled the "!" Symbol is shown on the screen to warn users
that the pump may operate without any manual intervention (pump resumes
with previous settings).
Do not use auto restart for more than 12 times in 24 hours. We recommend
remote control where a high number of starts is required.
If the pump is configured to network mode or analog mode it will respond to
remote commands at any time including immediately after power on. The "!"
Symbol is shown on the screen to warn users that the pump may operate
without any manual intervention (e.g. a remote setpoint could start the pump
with no key press required).
There are moving parts inside the pumphead. Before opening the tool-
unlockable guard or tool-unlockable track, ensure that the following safety
directions are followed:
1.
Ensure the pump is isolated from the mains power.
2.
Ensure that there is no pressure in the pipeline.
3.
If a tube failure has occurred, ensure that any fluid in the pumphead
has been allowed to drain to a suitable vessel, container or drain.
4.
Ensure that appropriate Personal Protective Equipment (PPE) is worn.
Primary operator protection from rotating parts of the pump is provided by
the pumphead safeguard. Note that safeguards differ, depending on the type
of pumphead. See the pumphead section of the manual.
12
m-630un-en-08
Содержание 630 U
Страница 21: ...1 2 m 630un en 08 21 ...
Страница 62: ...14 7 Network topology Star network Ring network 62 m 630un en 08 ...
Страница 126: ...23 5 Flow Sensor Reading The flow sensor value can be read via the Flow sensor reading screen 126 m 630un en 08 ...
Страница 128: ...Press SELECT to choose an option Press EXIT to return to the screen from where the MENU was called 128 m 630un en 08 ...
Страница 158: ...Press HOME to return and store the setting The input is now enabled in manual mode 158 m 630un en 08 ...
Страница 159: ...25 Help 25 1 Help Select Help from the main menu to access the help screens m 630un en 08 159 ...
Страница 179: ...Number Spare Description 8 TT0005 10mm 3 8in spanner 9 MRA3026A Track assembly m 630un en 08 179 ...