22.8
Pump parameters
Setting parameters
To set a parameter to a new value:
•
Type a value into the field or click on the check box (depending on the parameter type)
•
Click 'set' to store the new value or click 'refresh' to cancel the change
•
Up to 10 parameters are displayed per page. Use the < and > buttons to navigate between pages
The following parameters can be set by the user.
Param
No.
EDS Values
Read/Write
Default
Value
(EDS)
Parameter
Range
Comments
1
SetFlowCal
Write
15120
1 -
2147483647
Use to set the tube flow
calibration value. The flow
calibration value is set in
μL. For more information
on flow calibration
see
2
SetSpeed
Write
100
1 - 2200
Speed is set in Deci RPM.
Max speed depends on
head type.See
.
3
SetSpeedLimit
Write
2200
1 - 2200
Speed is set in Deci RPM.
Max speed depends on
head type. See
.
4
SetFailsafeSpeed
Write
100
1 - 2200
If the failsafe is enabled
the pump will run
continuously at the
selected speed in the
event of a
communications loss.
m-630un-en-08
101
Содержание 630 U
Страница 21: ...1 2 m 630un en 08 21 ...
Страница 62: ...14 7 Network topology Star network Ring network 62 m 630un en 08 ...
Страница 126: ...23 5 Flow Sensor Reading The flow sensor value can be read via the Flow sensor reading screen 126 m 630un en 08 ...
Страница 128: ...Press SELECT to choose an option Press EXIT to return to the screen from where the MENU was called 128 m 630un en 08 ...
Страница 158: ...Press HOME to return and store the setting The input is now enabled in manual mode 158 m 630un en 08 ...
Страница 159: ...25 Help 25 1 Help Select Help from the main menu to access the help screens m 630un en 08 159 ...
Страница 179: ...Number Spare Description 8 TT0005 10mm 3 8in spanner 9 MRA3026A Track assembly m 630un en 08 179 ...