Figure C–1: Recommended temperature sensor locations
A
B
C
D
E
F
1
2
3
4
5
6
7
8
TS
TS
TS
TS
TS
TS
Temperature sensor
Figure C–2: Minimum sample compartment temperature specifications
0
5
10
15
20
25
30
4
8
12
16
20
24
28
32
36
40
Ambient temperature (°C)
Maximum compartment temperature
Minimum compartment temperature
Minimum sustainable compartment
set point temperature
Sa
mple m
ana
ger
compa
rtmen
t
tem
p
er
ature
(°C)
+4 °C
-2 °C