Item
Specification
Number of sample
plates
Any two of the following Waters-certified plates:
• 96 and 384 microtiter plates
• 48-position, 2.00-mL vial plates
• 48-position, 0.65-mL micro-centrifuge tube plates
• 24-position, 1.50-mL micro-centrifuge tube plates
For more information, see the
Waters Sample Vials and Accessories
Brochure
, part number 720001818 or visit the
and
Maximum sample
capacity
768 in two, 384-well Waters-certified plates, or 96 in 2-mL vial holders.
Four additional positions for dilution functions.
For more information, see the
Waters Sample Vials and Accessories
Brochure
, part number 720001818 or visit the
and
Sample
compartment
temperature range
Between 4 and 40 °C, settable in increments of 0.1 °C; maintains 19 °C
below ambient with a tolerance range of between -2.0 and +4.0 °C
• At a set point of 4 °C with ambient temperature <23 °C and humidity
<80%, maintains a sample temperature of 2 to 8 °C.
• At ambient temperatures >23°C and humidity <80%, maintains an
average sample temperature of 18 °C below ambient, ±3.0 °C.
Minimum sample
compartment
temperature
specifications
See the graph that follows this table for the achievable sample
compartment temperature and expected variation at various sample
temperatures.
The graph shows achievable sample compartment temperature and
expected variation at various sample temperatures.
Recommended
locations for
temperature
sensors
See the diagram that follows this table for the recommended locations for
temperature sensors on the sample tray when validating specifications.
Temperature
accuracy
±0.5 °C at sensor
Temperature
stability
±1.0 °C (at the sensor with sample compartment door closed)
Injection needle
wash
Integral, active, and programmable
Minimum sample
required
3 µL, residual, using total recovery 2-mL vials (zero offset)
August 8, 2016, 715005049 Rev. C
Page 108