Emptying the roughing pump exhaust liquid trap bottle
December 6, 2012, 71500126802IVD Rev. C
67
2. Unscrew and remove the roughing pump liquid trap bottle cap.
3. Dispose of the waste liquid in accordance with local environmental
regulations.
4. Fit and tighten the roughing pump liquid trap bottle cap.
5. Ensure that the roughing pump liquid trap bottle is secured in the
upright position.
6. Open the source isolation valve (see
7. Use snoop (or equivalent) leak detector liquid to ensure that there are no
leaks at the liquid trap bottle cap.
Warning:
The waste liquid can be contaminated with
biohazardous and/or toxic materials. Ensure that it is
correctly disposed of according to local environmental
regulations.
Caution:
To avoid damage to the instrument, snoop (or
equivalent) leak detector liquid must be used only for the
purpose described in the following step; it must not be used on
any other part of the instrument.