4 Maintenance Procedures
60
December 6, 2012, 71500126802IVD Rev. C
Operating the source isolation valve
You must close the source isolation valve to isolate the source from the
instrument vacuum system for certain maintenance procedures.
Required material:
Chemical-resistant, powder-free gloves
To close the source isolation valve before starting a maintenance procedure
1. Prepare the instrument for working on the source (see
).
2. Open the instrument’s access door.
3. Open the source enclosure door by releasing both spring-clips and
lowering the door towards you.
Warning:
The source components can be contaminated with
biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
Warning:
To avoid electric shock, ensure that the instrument is
suitably prepared before commencing this procedure.
Warning:
The source can be hot. To avoid burn injuries, take
great care while working with the instrument’s access door open.
Warning:
To avoid puncture wounds, take great care while
working with the source enclosure door open if an ESI probe is
fitted (the probe tip is sharp).
Caution:
Do not apply any downward force to the source
enclosure door while the door is open.
Содержание TQ
Страница 34: ...1 Waters TQ Detector 34 December 6 2012 71500126802IVD Rev C...
Страница 144: ...4 Maintenance Procedures 144 December 6 2012 71500126802IVD Rev C...
Страница 156: ...A Safety Advisories 156 December 6 2012 71500126802IVD Rev C...
Страница 188: ...B External Connections 188 December 6 2012 71500126802IVD Rev C...
Страница 192: ...C Materials of Construction and Compliant Solvents 192 December 6 2012 71500126802IVD Rev C...