4 Maintenance Procedures
126
December 6, 2012, 71500126802IVD Rev. C
4. Use the O-ring removal kit to carefully remove the seal from the source
enclosure door.
5. Use the screwdriver to remove the 4 screws securing the glass-retaining
plate to the source enclosure door.
6. Remove the plate from the source enclosure door.
7. Remove the glass from the source enclosure door.
8. Use the O-ring removal kit to carefully remove the door glass seal from
the source enclosure door.
9. Dispose of all the seals in accordance with local environmental
regulations.
Warning:
The seals can be contaminated with biohazardous
and/or toxic materials. Ensure that they are correctly
disposed of according to local environmental regulations.
RP00019
Source enclosure
door seal
Source enclosure
glass seal
Door glass retaining
screw