Sterilization cycles
Load type
Placement
Pouched items
On trays allowing adequate space in-between
bags. Ensure that packs do not touch the walls of
the chamber. Place sterilization pouched items
with the paper side facing up.
Empty containers or non-
perforated trays
Upside down to prevent accumulation of water
Items made from different
materials (stainless steel,
carbon steel, aluminium, etc.)
On separate trays or wrapped/pouched
Instruments manufactured
from carbon steel
Place paper among them and the trays to avoid
rusty spots
PARTIAL LOAD
If the chamber is just partially loaded, place the load in such a way
that the space in-between the trays is maximized. Spread items
evenly on multiple trays. Below is a example with three trays.
Prepare the sterilizer
WARNINGS
Notice
: use only distilled or demineralized water (see "Feed water
specifications (EN 13060)" on page 109 for technical requirements).
Do not add any chemical / additive to the water.