Sterilizer tests
Lyla | Instructions for Use | RIN-210 ENG Rev02 | 15/09/2022 | © W&H Sterilization Srl
41
1
Empty the sterilizer chamber to ensure no load is present. Remove all trays from
the sterilization chamber, except the lowest one.
2
Place the Bowie and Dick test pack in the center of a tray in the lowest rack
position and close the chamber door.
3
On the homepage, tap
Helix-B&D test
.
4
To set the duration of the Plateau/Sterilization phase and other settings, tap
.
5
Tap
and enter your credentials if required: the chamber door locks.
6
Wait until the end of the test and tap
OPEN
: the chamber door unlocks.
7
Enter your credentials if required.
8
Open the chamber door, extract the tray using the tray holder and take the test
pack. The test pack can be wet outside.
CAUTION!
Risk of burns. The test pack is very hot at the end of the cycle. Wear
appropriate PPE (e.g. gloves).
9
Remove the indicator sheet from the center of the test pack and check the change
in colour. See "Interpret the test result" on the next page
Содержание Lyla
Страница 1: ...RIN 210 17 RIN 210 22 Instructions for Use RIN 210 ENG Rev02...
Страница 2: ......