Diagnostics
Lyla | Instructions for Use | RIN-210 ENG Rev02 | 15/09/2022 | © W&H Sterilization Srl
85
MESSAGES AND ALERTS
Notice
: for any error not listed in this table, call technical service.
Message/Alert
Description
Action
Fill clean water tank.
There is not enough water in the tank to perform a
cycle.
Fill the water tank as requested.
Drain used water tank.
The used water tank is full.
Drain the water tank as requested.
Please close the door.
The door must be locked, but you didn't close it.
Close the door so it can be locked.
Non-conform water
The clean water quality is bad (conductivity
between 15 and 50 μS/cm).
You may run a cycle but the water must be replaced soon, otherwise the unit will
automatically lock-out to prevent damage.
Unacceptable water
The clean water quality is very bad (conductivity
more than 50 μS/cm).
Running a cycle is inhibited to prevent damage. Replace the clean water.
Door Gasket must be replaced in
... cycles.
Have you ordered the Door
Gasket?
These are pre-alerts advising that one of the
consumables has to be replaced within a small
number of cycles.
Tap
if you have the consumable available for replacement.
Tap
if you don't have the consumable in stock and must order one. In this case, the
pre-alert will appear again after some cycles.
See "Maintenance" on page 60.
Bact. Filter must be replaced in
... cycles.
Have you ordered the Bact.
filter?
Dust Filter must be replaced in
... cycles.
Have you ordered the Dust
filter?
4000 cycle maintenance must
be performed in ... cycles.
Have you already booked the
4000 cycle service?
This pre-alert advises you that the target of 4000
cycles is close and the relevant maintenance step
is to be scheduled.
Contact your Technical Service.