Diagnostics
Troubleshooting
MANAGING ERRORS
If during a sterilization cycle an error occurs do the following:
1
Wait until the end of the reset phase.
CAUTION!
Do not switch off the sterilizer during the reset
phase: it takes some minutes to reset the system and reach
safe conditions in the sterilizer chamber.
2
When the
OPEN
button appears, tap it to unlock the door.
3
Confirm the opening of the door.
Notice
: water could be present in the chamber when opening the
door: prevent spilling (e.g., place a towel below the chamber door).
VIEW AND SAVE THE ERROR LOG
1
On the homepage, tap
>
>
>
: a list of the last
errors appears.
2
Tap
to save the list in the USB pen drive.
ERROR PAGE
During the sterilization cycle, the sterilizer is continuously monitored
by a control system. If an anomaly is detected, the cycle is aborted
automatically, and the sterilizer starts a reset phase.
The following page appears:
Part
Description
A
Current
sterilization cycle
B
Error number, See
C
Warning
messages.
D
Open button that
appears after the
reset phase had
finished.
WARNING MESSAGES
Message
Description
Load not
sterile
The load is not sterile.
WARNING! Do not use items on patients!
Drying
interrupted
The load might be wet.
WARNING! Wet items are for immediate use only!