Maintenance
LISA | Instructions for Use | VA131 2019 ENG Rev06 | 16/04/2020 | © 2020 W&H Sterilization Srl
89
PREPARE THE STERILIZER FOR CLEANING THE WATER TANKS
A video animation is available on https://video.wh.com/en_global/dental-faq.
Proceed as follows:
1
Switch OFF the sterilizer and disconnect the mains cable.
2
Completely drain both water tanks (see "Draining the used and clean water tank"
on page 98).
3
To drain the detergent solution during the subsequent cleaning, leave the drain
tube attached to the drain port of the tank you want to clean.