Maintenance
80
LISA | Instructions for Use | VA131 2019 ENG Rev06 | 16/04/2020 | © 2020 W&H Sterilization Srl
Monthly or 50-cycle maintenance
CLEANING THE DOOR GASKET AND THE CHAMBER FACE SIDE
Proceed as follows:
1
Clean the seal seat and the chamber face side with a damp, lint-free cloth
moistened with clean water.
Notice:
Do not use abrasive products, cutting tools or sharp objects.
If you use a detergent solution, be careful not to get in contact with the plastic
body of the front cover.
2
Rinse with clean water.
Note: when the seal is new it might be necessary to hold the door gently closed at the
sterilization start.