Sterilizer tests
50
LISA | Instructions for Use | VA131 2019 ENG Rev06 | 16/04/2020 | © 2020 W&H Sterilization Srl
3
On the homepage tap Helix-B&D test.
4
To set the duration of the Plateau/Sterilization phase and other settings, tap
.
5
Tap
and enter your credentials if required: the chamber door locks.
6
Wait until the end of the test and tap OPEN: the chamber door unlocks.
7
Enter your credentials if required.
7
Open the chamber door, extract the tray using the tray holder and take the test
pack. The test pack can be wet outside.
CAUTION! Risk of burns. The test pack is very hot at the end of the cycle.
Wear appropriate PPE (e.g. gloves).
8
Remove the indicator sheet from the center of the test pack and check the
change in colour. See "Interpret the test result" on the next page