Technical data
LISA | Instructions for Use | VA131 2019 ENG Rev06 | 16/04/2020 | © 2020 W&H Sterilization Srl
117
STANDARD STERILIZATION CYCLES AVAILABLE
There are four sterilization cycles available, all of them complying with the European Standard EN13060:
n
three B-type cycles
n
one S-type cycle
Cycle
type
Cycle name
Purpose
B
B Universal
134
For all your general items like hand instruments, handpieces, forceps, etc.
B Extended
134
Features an extended sterilization time, if this is required for your load or mandated in your country.
B Universal
121
For all items that cannot withstand the high temperatures of the 134 cycles, such as textiles and plastics.
S
S Fast 134
For fast processing of unwrapped instruments only, including dental turbines and handpieces, solid, and hollow B (simple hollow items). It is not suitable
for textiles, porous or bagged/wrapped items.
After the cycle is completed, the sterilizer could perform two drainage periods of about 30 seconds each.
The instruments sterilized with this cycle cannot be stored: they must be used immediately after being sterilized.
COMMON STERILIZATION CYCLE DATA
Sterilization cycles
B Universal 134
B Extended 134
B Universal 121
S Fast 134
Sterilization temperature
134 °C (273 °F)
134 °C (273 °F)
121 °C (250 °F)
134 °C (273 °F )
Sterilization pressure
3.03 bar
2.03 bar (g)
3.03 bar
2.03 bar (g)
2.04 bar
1.04 bar (g)
3.03 bar
2.03 bar (g)
Duration of the Plateau/Sterilization phase
4'
18' 30"
20' 30"
3' 30"
Duration of the drying phase (ECO DRY plus mode)
5'–16'
5'–16'
6'–24'
1'–4'
Содержание Lisa VA131-17
Страница 1: ...VA131 17 VA131 22 Instructions for Use VA131 2019 ENG Rev06 ...
Страница 2: ......
Страница 139: ......