Version: 1
Issued : February, 2016

INSTRUCTION MANUAL

VWR

®

 PURANITY TU+ 

European Catalogue Number(s)

171-1231
171-1232 
171-1233 
171-1234 
171-1235 
171-1236 
171-1237 
171-1238 
171-1239 

Puranity TU3+
Puranity TU3 UV+
Puranity TU3 UV/UF+
Puranity TU6+
Puranity TU6 UV+
Puranity TU6 UV/UF+
Puranity TU12+
Puranity TU 12 UV+
Puranity TU 12 UV/UF+

Содержание Puranity TU 12 UV+

Страница 1: ...TU European Catalogue Number s 171 1231 171 1232 171 1233 171 1234 171 1235 171 1236 171 1237 171 1238 171 1239 Puranity TU3 Puranity TU3 UV Puranity TU3 UV UF Puranity TU6 Puranity TU6 UV Puranity T...

Страница 2: ...Legal Address of manufacture VWR International bvba Researchpark Haasrode 2020 Geldenaaksebaan 464 B 3001 Leuven 32 16 385011 http be vwr com Country of origin Sweden...

Страница 3: ...ption and cost repair replace or refund the purchase price of this product to the customer provided it is returned during the warranty period This warranty does not apply if the product has been damag...

Страница 4: ...ch if not avoided will result in death or serious injuries NOTE Is used for applicational hints and useful information Risk of electric shock Electrical work on the system is only to be carried out by...

Страница 5: ...on the back of the unit on the nameplate The catalog number Compliance with local laws and regulations The customer is responsible for applying for and obtaining the necessary regulatory approvals or...

Страница 6: ...vi Puranity TU VWR...

Страница 7: ...ections of the Puranity TU 3 6 system 7 21 Connections of the Puranity TU 12 system 7 23 Connections of the pure water tank Puranity TU 12 7 25 Bring your Puranity TU system into operation 7 26 Connec...

Страница 8: ...placing the RO pretreatment 13 52 Cleaning 13 53 Changing the ultrafilter 13 56 UV reactor assembly 13 58 Replacing the UV lamp 13 59 Replacing the 0 2 m final filter 13 61 Autoclaving the 0 2 m final...

Страница 9: ...l possibly occur during transport Lifting and carrying the VWR Ultrapure Water System e g to the installation location should be carried out by two people Examination on receipt Check the completeness...

Страница 10: ...n shipment If possible use the original box and packaging material When these are no longer available then Protect the system from shock by packing it in bubble wrap and or packaging foam and a strong...

Страница 11: ...location should be carried out by two people To do this lift the system in tandem at the two corner points beneath the bottom plate The CE mark is nullified if you make any structural changes to the...

Страница 12: ...t any supply of feedwater The manufacturer will not accept any liability should this occur The system must be disinfected or rinsed after an extended down time The cleaning procedure is described unde...

Страница 13: ...always wear appropriate safety gloves to prevent skin damage or burns Wear safety gloves when changing the UV lamp in order to prevent that your skin comes in contact with the UV lamp glass Wear safet...

Страница 14: ...tions 6 Puranity TU VWR Chapter 2 Proceed as follows when the system has a defect Switch the system off and unplug the system from power outlet Shut off the feedwater supply Contact your local service...

Страница 15: ...2 for a 6L h system 171 1143 including 1 x assembly kit 171 1140 consisting of 1 x Final final filter sterile filter 0 2 m 171 1105 1 x Sterile vent air filter 171 1166 1 x Feedwater connecting kit 17...

Страница 16: ...ith 30L capacity or 171 1170 60L capacity 171 1171 1 x Ultrapure cartridge 171 1141 1 x RO pretreatment cartridge 171 1165 including 1 x assembly kit 171 1185 consisting of 1 x Final final filter ster...

Страница 17: ...torage tanks for Puranity TU 12 system Available Puranity TU 12 systems Cat No System Capacity 171 1170 Pure water tank 30 liter 171 1171 Pure water tank 60 liter Cat No System 171 1237 Puranity TU 12...

Страница 18: ...Extend of delivery 10 Puranity TU VWR Chapter 3...

Страница 19: ...on of water into the water categories mentioned in the standards of ASTM 11 01 and ASTM 11 02 The system must not be operated outside of the specifications as described in the operating manual In part...

Страница 20: ...Intended use of the device 12 Puranity TU VWR Chapter 4...

Страница 21: ...05 ppm pH range 4 11 Temperature 2 C 35 C Pressure 1 6 bar 14 87 PSI at a pressure 6 bar 87 PSI a pressure reducer must be installed upstream of the system Product water quality ASTM Type I on dispens...

Страница 22: ...Resistance reference temperature 25 C M cm 15 10 15 10 15 10 Retention rate for bacteria and particles 99 99 99 Tank capacity Puranity TU 3 6 Liter 5 5 5 External tank capacity Puranity TU 12 Liter u...

Страница 23: ...01 inch Depth 300 mm 11 81 inch Base area 277 x 372 mm 10 91 x 14 65 inch Weight including ultrapure cartridge and RO pretreatment cartridge 28 kg 61 74 lbs Dimensions and weight 30L and 60L pure wat...

Страница 24: ...ameter Tank Flow tubing only Puranity TU 12 Tank back flow tubing only Puranity TU 12 Tube 1 4 outer diameter Tank overflow tubing only Puranity TU 3 6 Tube 1 4 outer diameter Dispenser back flow tubi...

Страница 25: ...anity TU systems and pure water tank Space on left and right from the side of the system at least 300 mm 11 81 inch Space to the back of the system at least 200 mm 7 87 inch Top space at least 400 mm...

Страница 26: ...18 Puranity TU VWR Chapter 5 Ventilation requirements no special requirements no special requirements Degree of pollution 2 2 Ambient conditions During operation Storage Airborne sound emission Sound...

Страница 27: ...ter not less than 1 bar 14 PSI and not greater than 6 bar 87 PSI CAUTION The feedwater pressure must not be allowed to go above 6 bar Install an additional pressure reducer when the feedwater pressure...

Страница 28: ...r Europe A check valve is recommended in the feedwater line to prevent back flow of feedwater from the water system An electric socket with protective connection must be available for connection of th...

Страница 29: ...VWR Puranity TU 21 Installation Connections of the Puranity TU 3 6 system 12 11 10 9 8 7 6 1 2 3 4 5...

Страница 30: ...or additional hand dispenser 6 Connector potential free contact 5 pins 7 Power supply connector 24 V DC 8 RS232 printer connection 9 1 4 thread connector for 0 2 m final filter 10 Quick connectors for...

Страница 31: ...VWR Puranity TU 23 Installation Chapter 7 Connections of the Puranity TU 12 system GHQWLDO 12 11 1 10 9 2 6 5 8 7 4 3...

Страница 32: ...for additional hand dispenser 7 Dispenser flow connector outer diameter 1 4 tube for additional hand dispenser 8 Optional printer connector 9 Power supply connector 24V DC 10 Level control connector...

Страница 33: ...13 Tank flow connector for outer diameter 1 4 tube 14 Tank back flow connector for outer diameter 1 4 tube 15 Level control connector 16 Connector for optional sterile overflow 17 Connector for option...

Страница 34: ...nt kit is available for wall mounting purchasing separately See Section Wall mounting on page 31 2 Remove the cover from the ultrapure cartridge by pressing the snapper and pull the cover off towards...

Страница 35: ...cartridge incorrectly 5 a Use the R3 4 feedwater connecting kit supplied with the system to connect the Puranity TU system with the feedwater tap Therefore push the R3 4 union nut over the outer diam...

Страница 36: ...e Puranity TU 12 panel see Connections of the Puranity TU 12 system on page 23 Route the outer diameter 1 4 tube with a free gravity fall to the drain see Illustration of drain on page 30 8 Connect th...

Страница 37: ...e installation guide If not use a AD 8mm 0 31 inch tube to connect the tank overflow to drain as described in this chapter A tank vent filter has to be ordered separately see Accessories on page 67 St...

Страница 38: ...oncentrate outlet tube NOTE Shorten the rinsing water and concentrate outlet tube to the required length and route it to an atmospherically vented drain The tubes that run from the Puranity TU system...

Страница 39: ...the wall mounting brackets are only suitable for attaching the wall mounting brackets to a concrete wall or a solid masonry wall Step Action Figure 1 Use the wall mounting brackets to mark the anchori...

Страница 40: ...he suitable plug and the power cable do not get wet Mount the power pack with dry hands Risk of an electrical shock Step Action Figure 1 Remove the protective foil on the back of the universal holder...

Страница 41: ...s been mounted 4 Now connect the power supply unit to the Power supply connection on the right side of the Puranity TU 3 6 or Puranity TU 12 system Next plug the power supply to a grounded 100 240V 50...

Страница 42: ...ion panel of the system c Cut from the outer diameter 1 4 tube two tubes with a length of 80 mm 3 2 inch and attach the tubes into the t adapter and the connection panel Attach the ball valve with a t...

Страница 43: ...g flow chart describes the Puranity TU Systems with full equipment ultrafilter UV lamp Depending on your Puranity TU system configuration the UV lamp or ultrafilter are inapplicable The flow chart dir...

Страница 44: ...Flow chart 36 Puranity TU VWR Chapter 8...

Страница 45: ...er solenoid valve is closed by a level switch and prevents that no more water flows into the system After open the dispensing valve the pure water from the tank flows through the UV bulb only systems...

Страница 46: ...How the Puranity TU system functions 38 Puranity TU VWR Chapter 9...

Страница 47: ...21 or shown in the bottom diagram 1 Concentrate connector for outer diameter 1 4 tube 2 Feedwater connector for outer diameter 1 4 tube 3 Tank overflow connector for outer diameter 1 4 tube 4 Tank bac...

Страница 48: ...re you start with dispensing water from the sampling tap Wait until the tank filling has been completed This could take approx 1 5 to 3 hours with a Puranity TU 3 6 system With a Puranity TU 12 system...

Страница 49: ...is shown in the display B Second line shows the currently running measurement mode in S cm M cm and C C Third line shows the Limit value Cond Temp and Status Sensor The Limit value Cond LED s will rep...

Страница 50: ...VWR Chapter 11 D Menu button switches the Menu to the next submenu E Down button decreases adjustable values F Up Quit button increases adjustable values quits acoustic alarms G Enter button confirms...

Страница 51: ...lue Mono dual measurement mode NOTE This following combinations are possible in the Mono dual measurement mode adjusted resistance value M cm c and ct adjusted resistance value S cm c and ct Mono meas...

Страница 52: ...LED lights up The first line shows the selected unit of measurement resistance value factory adjustments in M cm 2 Use the Up Quit and down button to choose the resistance value for measuring mode S c...

Страница 53: ...ess the Menu button twice to enable the limiting value for the conductivity to be set red LED lights up 2 Use the Up quit and down button to increase or reduce the value With the Enter button you swit...

Страница 54: ...ree times to enable the limiting value for the temperature to be set red LED lights up 2 Use the Up quit and down button to increase or reduce the value With the Enter button you switch to the next di...

Страница 55: ...ionally after approx 5 minutes an acoustic signal will sound Error Lights up LED in second line and acoustic signal Limiting value of conductivity is reached Limiting value of temperature is reached L...

Страница 56: ...atically the conductivity is scaled to three significant places Output is made 1x per hour When the system is in operation mode a press on the down button triggers a measured value output at the seria...

Страница 57: ...in an easily accessible location for anyone who is using or servicing the system Calibration of the conductivity may only be performed by a service technician authorized by the manufacturer Cleaning...

Страница 58: ...er occurs first Bacteria growth may occur in the resin when the system has been in use over an extended period Sterile filter 0 2 m F2 171 1105 12 months Or the flow rate is markedly slower Ultrafiltr...

Страница 59: ...cartridge and insert the cartridge into the system 6 Insert the quick connectors into the connecting points on the ultrapure cartridge When you hear an audible click you can be sure that the quick con...

Страница 60: ...tors on the cover of the RO pretreatment cartridge and remove the used cartridge from the system 4 Locate the new RO pretreatment cartridge and insert the cartridge into the system 5 Insert the quick...

Страница 61: ...aning Cleaning Solution 1 syringe 171 1124 NOTE To ensure effective cleaning of your system the internal 5L or external tank must be filled completely with water There is no need to worry should you n...

Страница 62: ...ave been inserted correctly 3 Always wear protective gloves and goggles when handling cleaning solution a Switch the system on connect main power supply open the feedwater tap and wait until the inter...

Страница 63: ...ain the content of water into the drain by open the dispensing valve on the tank Open the feedwater tap and wait until the tank is filled for run one tank filling through the sampling tap c Switch the...

Страница 64: ...nt on page 52 c Use the Phillips screwdriver to unscrew the screws from the bracket for the UV lamp and pull the UV assembly out toward the front d Unscrew the threaded connection screws by hand on th...

Страница 65: ...in step 3b and screw the fitting into the same position that it was on the old ultrafilter Use the 17mm 0 67 inch open end wrench for this c Screw on the measuring cell and insert the tubes into the t...

Страница 66: ...ance 58 Puranity TU VWR Chapter 13 UV reactor assembly Union nut Seal ring O ring Immersion tube UV lamp casing UV lamp Immersion tube Stopper Stopper These stoppers are located in the UV lamp casing...

Страница 67: ...1 Switch the Puranity TU system off Unplug from main power supply 2 Remove the cover for the ultrapure cartridge 3 a Take the ultrapure cartridge out of the system see Replacing the ultrapure cartrid...

Страница 68: ...e provided for it on the top of the union nut figure 1 and 2 The O ring fits in the bottom groove in the union nut figure 1 If these items do not fit exactly in the grooves and you restart the system...

Страница 69: ...tridge too see Replacing the RO pretreatment on page 52 8 Switch the system on connect main power supply Step Action Figure 1 Rotate the exhausted or clogged 0 2 m final filter in a clockwise position...

Страница 70: ...n a steam autoclave at 121C for 30 minutes You can repeat the sterilization process for the sterile filter up to 5 times When sterilization is completed screw the 0 2 m final filter back into the 1 4...

Страница 71: ...also your responsibility to decontaminate the equipment in case of biological chemical and or radiological contamination so as to protect from health hazards the persons involved in the disposal and r...

Страница 72: ...Equipment disposal 64 Puranity TU VWR Chapter 14...

Страница 73: ...ridge blocked Open the feedwater supply line Replace connections Increase feedwater pressure Replace the sterile filter see Replacing the 0 2 m final filter on page 61 Install new RO pretreatment cart...

Страница 74: ...Replace with new ultrapure cartridge Check and reset the limiting value Error message Red LED in second line is light up Limit value Temp Temperature in system is too high Limiting value is set too lo...

Страница 75: ...odule 171 1106 Sterile filter 0 2 m 171 1105 RO pretreatment cartridge for Puranity TU 3 Puranity TU 6 Puranity TU 12 171 1142 171 1143 171 1165 Sterile vent air filter 171 1166 Designation Cat No Pri...

Страница 76: ...Consumables 68 Puranity TU VWR Chapter 16...

Страница 77: ...VWR Puranity TU 69 Terminal assignment...

Страница 78: ...Terminal assignment 70 Puranity TU VWR Chapter 17...

Страница 79: ...ncentrate flow contact your VWR service department Customer address Location System type Serial no Year made Date Ultrapure water resistance M xcm Temperature C Last change of RO pretreatment cartridg...

Страница 80: ...Maintenance record 72 Puranity TU VWR Chapter 18...