Warning
The Sterilization Cycle heats the incubator interior surfaces to
140°C. Contact with any surface inside the outer door during this cycle
may result in burns.
s
1. Empty the water from the humidity pan and place the pan back into
the incubator. Remove any samples, instruments, dishes, etc. from the
chamber.
2. Press and hold the large white sterilization cycle button on the right
side of the unit (Figure 5-4) for approximately 3 seconds until the
LED lights.
3. ENTER the access code, if applicable. An access code is recommended
to prevent accidental cycle initiation.
4.
Pre Sterilization with T/C CO
2
sensor
- The display toggles between
“REMOVE HEPAs”, “REMOVE WATER” and “PRESS ENTER”. If
ENTER is not pressed within 1 minute, the display returns to
“SYSTEM OK”.
5. Remove the HEPA filter,
air sample filter and access
port filter assembly (Figure
5-5). Tubing can remain in
the unit during the
sterilization cycle. In
addition, remove the
temp/RH recorder probe
and IR sensor, if applicable.
Discard the filters.
6.
Pre Sterilization with IR CO
2
sensor
- If an IR sensor is connected,
the display toggles between “POWER OFF” and “REMOVE IR”. If
power is not turned off within 1 minute, the display returns to
“SYSTEM OK”.
Sterilization Cycle
Figure 5-4.
Sterilization Button Location
Air Sample
Filter
Access
Port Filter
and stopper
assembly
Temp/RH
recorder probe
HEPA
Filter
IR sensor
(behind plate)
Temp sensor
(inserted into
HEPA collar from
behind top duct)
Figure 5-5.
Filter Locations
RUN
SET
CAL
CON
MODE
HEAT
INJ
T (°C)
RH (%)
CO (%)
2
SCROLL FOR PARAMETERS
ENTER
Sterilization
Button
Air Jacketed CO2 Incubator
5-7
VWR International
Section 5
Routine Maintenance