The low CO
2
alarm limit is the deviation from the CO
2
setpoint which
will cause a low CO
2
alarm. The setpoint is variable from 0.5% CO
2
below setpoint to 5.0% CO
2
below setpoint. The factory setting is 1.0%
CO
2
below setpoint. A minus sign (-) in the display indicates that the
alarm setting is below the setpoint.
1. Press the MODE key until the CON indicator lights.
2. Press the right arrow until CO2 LO -X.X shows in the message
display.
3. Press up/down arrow to change the low CO
2
alarm limit.
4. Press ENTER to save the low CO
2
alarm limit.
5. Press the MODE key to return to RUN mode or right/left arrow to go
to next/previous parameter.
The high CO
2
alarm limit is the deviation from the CO
2
setpoint which
will cause a high CO
2
alarm. The setpoint is variable from 0.5% CO
2
above setpoint to 5.0% CO
2
above setpoint. The factory setting is 1.0%
CO
2
above setpoint.
1. Press the MODE key until the CON indicator lights.
2. Press the right arrow until CO2 HI X.X shows in the message display.
3. Press up/down arrow to change the high CO
2
alarm limit.
4. Press ENTER to save the high CO
2
alarm limit.
5. Press the MODE key to return to RUN mode, or right/left arrow to
go to next/previous parameter.
3-4
Air Jacketed CO2 Incubator
VWR International
Section 3
Configuration
Set Low CO
2
Alarm
Limit (tracking alarm)
Set High CO
2
Alarm
Limit (tracking alarm)