![VWR 10810-902 Скачать руководство пользователя страница 30](http://html.mh-extra.com/html/vwr/10810-902/10810-902_operating-instructions-manual_1057410030.webp)
Air Jacketed CO2 Incubator
3-1
VWR International
Section 3
Configuration
Several features available in Configuration mode allow custom setup of the
incubator. These features are listed with descriptions below. All features
may not be necessary in all applications, but are available if needed. To
enter Configuration mode, press the MODE key until the CON indicator
lights. Press the right and/or left arrow until the appropriate parameter
appears in the message display. See Chart 3-1 for more detail.
The audible alarm can be turned on or off. The factory setting is ON.
1. Press the MODE key until the CON indicator lights.
2. Press the right arrow until AUDIBLE XXX shows in the message
display.
3. Press the up/down arrow to toggle AUDIBLE ON/OFF.
4. Press ENTER to save the setting.
5. Press the MODE key to return to RUN mode or right/left arrow to go
to next/previous parameter.
When the REPLACE HEPA reminder displays and the visual alarm
flashes, the specified time has elapsed and the HEPA filter should be
replaced. To clear the display and reset the timer after replacing the HEPA
filter with a new one, follow the steps below.
1. Press the MODE key until the CON indicator lights.
2. Press the right arrow until NEW HEPA shows in the message display.
3. Press ENTER to restart the timer and clear the REPLACE HEPA
alarm.
4. Press the MODE key to return to RUN mode.
New HEPA Filter
Turn Audible Alarm