5-2
Air Jacketed CO2 Incubator
VWR International
Section 5
Routine Maintenance
1. Remove the shelves, access port filter assembly, HEPA filter, air sample
filter and tubing, and the left and right duct sheets. Discard the HEPA
filters and the access port filter assembly. See Figure 5-1.
2. Wash the air sample filter tubing, shelves, and ducts with disinfectant,
then rinse with sterile water. Optionally, the shelves and ducts may be
autoclaved.
3. Wash the inner door gasket with disinfectant. This gasket may be
removed to be cleaned, or replaced.
4. Wash the cabinet interior with disinfectant, starting at the top and
working down. Refer to the disinfectant directions for length of time
needed before rinsing. Wash the inner door both inside and out. The
cabinet and door must be rinsed with sterile water until the
disinfectant has been removed. After the cabinet has been rinsed, spray
with 70% alcohol.
5. Install the left and right ducts, inner door gasket, access port filter
assembly, and air sample filter and tubing, spraying each with 70%
alcohol.
6. Install a new HEPA filter.
7. Install the shelves and spray with 70% alcohol.
Clean the incubator exterior with a damp sponge or soft, well-wrung cloth
and mild detergent dissolved in water. Dry with a soft cloth.
Clean Cabinet
Exterior
Clean Chamber
Interior
Air Sample
Filter
and tubing
Access
Port Filter
and stopper
assembly
HEPA
Filter
Humidity Pan
Figure 5-1.
Component Locations