![VWR 10810-902 Скачать руководство пользователя страница 38](http://html.mh-extra.com/html/vwr/10810-902/10810-902_operating-instructions-manual_1057410038.webp)
Air Jacketed CO2 Incubator
4-1
VWR International
Section 4
Alarms
The VWR Air Jacketed Incubator alarm system is shown in the table
below. When an alarm is active, the message appears in the LED message
display. Pressing
disables the audible alarm for the ringback period.
However, the visual alarm continues until the incubator returns to a
normal condition. The alarms are momentary alarms only. When an alarm
condition occurs and then returns to normal, the incubator automatically
clears the alarm condition and the message display.
When multiple alarm conditions occur, active messages are shown in the
message display one at a time, updating at 5 second intervals. Pressing
during multiple alarms causes all active alarms to be silenced and to
ring back in 15 minutes.
The TEMP IS LOW alarm is disabled when the Temp set point is 10°C.
The CO
2
alarms are disabled when the CO
2
set point is 0.0%.
Description
Message
Delay
Ringback
Relay
No alarm condition exists
SYSTEM OK or CLASS 100
-----
-----
-----
Temp > Otemp Set point
SYS IN OTEMP
0 min.
15 min.
Yes
Air Temp Sensor Fault (See Section 4)
AIR SNSR ERR
0 min.
15 min.
No
Temperature Controller Failure (See Sect. 4)
TMP CTRL ERR
0 min.
15 min.
YES
CO2 Sensor Fault (See Section 4)
CO2 SNSR ERR
0 min.
15 min.
No
Replace HEPA filter reminder-set time expired
(See Section 3)
REPLACE HEPA
0 min.
-----
No
Inner Door is Open
DOOR OPEN
15 min.
15 min.
No
CO2 is higher than CO2 High Tracking Alarm
CO2 IS HIGH
15 min.
15 min.
Programmable
CO2 is lower than CO2 Low Tracking Alarm
CO2 IS LOW
15 min.
15 min.
Programmable
TEMP is lower TEMP Low Tracking Alarm
TEMP IS LOW
15 min.
15 min.
Programmable
Table 4-1.
Alarm and Description
- All alarm delays and ringback times are ±30 seconds -