![VWR 10810-902 Скачать руководство пользователя страница 3](http://html.mh-extra.com/html/vwr/10810-902/10810-902_operating-instructions-manual_1057410003.webp)
VWR International
ii
Air Jacketed CO2 Incubator
Preface
Contains Parts and Assemblies
Susceptible to Damage by
Electrostatic Discharge (ESD)
CAUTION
Important
Read this instruction manual. Failure to read, understand and follow the instructions in this manual
may result in damage to the unit, injury to operating personnel, and poor equipment performance.
s
Caution
All internal adjustments and maintenance must be performed by qualified service personnel.
s
Material in this manual is for information purposes only. The contents and the product it describes are subject
to change without notice. VWR International makes no representations or warranties with respect to this
manual. In no event shall VWR
*OUFSOBUJPOBM
be held liable for any damages, direct or incidental, arising out of
PS
related to the use of this manual.
Intended Use:
The VWR Air Jacketed CO2 Incubators are designed to maintain an optimal environment for the incubation of
tissue and cell samples. These models are designed to maintain temperature and carbon dioxide levels as set by
the operator as well as monitoring chamber relative humidity. CO2 is controlled using either a thermo-
conductivity cell, or an infra-red sensor.
The VWR Air Jacketed Incubators are approved for general purpose use only.
Non-intended Use:
The VWR Air Jacketed Incubators are not intended for use where electrical or physical contact with the patient
is established.
These incubators are not intended to be operated in potentially explosive environments and are not intended for
use with flammable materials.
The VWR Air Jacketed Incubators are not intended for use as a Microbiological Incubator (21 CFR 866.2540)
and are also not approved for use in assisted reproductive procedures for the incubation of ova and embryos (21
CFR 884.6120).
Warning
If the incubator is not used in the manner specified in this operating manual, the protection provided
by the equipment design may be impaired.
s
©2015 VWR International. All rights reserved.