1. Measure the CO
2
concentration in the chamber through the gas
sample port with a Fyrite or other independent instrument. Several
readings should be taken to ensure accuracy.
2. Press the MODE key until the CAL indicator lights.
3. Press the right arrow until “IR CAL XX.X” appears in the message
display.
4. Press the up/down arrow to adjust the display to match the
independent instrument reading.
5. Press ENTER to store the calibration.
6. Press the MODE key to return to RUN mode.
2-4
Air Jacketed CO2 Incubator
VWR International
Section 2
Calibration
Infrared CO
2
System
(continued)