63
Alternating voltage
Direct current
Alternating current
Range
Accuracy
Resolution
Frequenz range
400 µA
0,01 µA
±(0,6 % + 40)
45 Hz - 1 kHz
±(1,2 % + 40)
1 kHz - 10 kHz
4000 µA
0,1 µA
±(0,6 % + 40)
45 Hz - 1 kHz
±(1,2 % + 40)
1 kHz - 10 kHz
40 mA
0,001 mA
±(0,6 % + 40)
45 Hz - 1 kHz
±(1,2 % + 40)
1 kHz - 10 kHz
400 mA
0,01 mA
±(0,6 % + 40)
45 Hz - 1 kHz
±(1,2 % + 40)
1 kHz - 10 kHz
10 A
0,001 A
±(2 % + 40)
45 Hz - 1 kHz
±(4 % + 40)
1 kHz - 10 kHz
Overload protection: Fuses, measuring time limit >5 A: max. 10 s with 10 min break
Overload protection 1000 V
TrueRMS in the measuring range of 10 – 100%
Range
Accuracy
Resolution
400 µA
0,01 µA
4000µA
0,1 µA
40 mA
0,001 mA
400 mA
0,01 mA
10 A
±(1,5% + 20)
0,001 A
Overload protection: Fuses, measuring time limit >5 A: max. 10 s with 10 min break
Range
Accuracy
Resolution
Frequenz range
4 V
0,0001 V
±(0,5 % + 40)
45 Hz - 1kHz
±(1,2 % + 40)
1 kHz - 10 kHz
±(3 % + 40)
10 kHz - 20 kHz
±(4 % + 40)
20 kHz - 100 kHz
40 V
0,001 V
±(0,5 % + 40)
45 Hz - 1kHz
±(1,2 % + 40)
1 kHz - 10 kHz
±(3 % + 40)
10 kHz - 20 kHz
±(6 % + 40)
20 kHz - 100 kHz
400 V
0,01 V
±(0,5 % + 40)
45 Hz - 1kHz
±(1,2 % + 40)
1 kHz - 10 kHz
±(3 % + 40)
10 kHz - 20 kHz
Not specified
20 kHz - 100 kHz
1000 V
0,1 V
±(1,2 % + 40)
45 Hz - 1kHz
±(3 % + 40)
1 kHz - 5 kHz
±(6 % + 40)
5 kHz - 10 kHz
Not specified
10 kHz - 100 kHz
Overload protection 1000 V; Impedance: 10 MΩ
TrueRMS in the measuring range of 10 – 100%: Crest factor: max. 3.0 (at 750V max. 1.5)
±(0,3% +10)
±(0,5% +10)