50
Conduct a zero reconciliation without conductor and with the current probe closed. For this,
briefly push the button “REL”. A beep and the delta symbol Δ in the display confirm the zero
reconciliation. To deactivate this function, push the button “REL” again. The Delta symbol
goes out. You are then back in normal measuring mode without zero reconciliation.
- Push the current probe opening lever (3) and open
the current probe this way.
- Observe the correct polarity of the current probe for
the direct current measurement. When connected
correctly, the cable must be routed to the consumer
through the current probe from the front when com-
ing from the power source (+).
- Reach around the individual conductor to be meas-
ured and close the current probe again. Position the
conductor centrally with the opening at the probe.
- The measured value is indicated on the display.
When a minus “-” appears in front of the meas-
ured value when measuring direct current,
the current has the opposite direction (or the
polarity of the current probe is swapped).
- Remove the current probe from the object to be
measured after completion of the measurement and
switch off the device. Turn the dial switch to “OFF”.