Page 13 of 65
3 vLoc3-DM Receiver Functions and Operations
Transmitter Link
- (if installed) (not available for the Loc-150Tx Transmitter)
This is a factory-fit option that must be
requested at the time of ordering. The Loc3 series transmitters can be remotely operated from the vLoc3 series receivers. This
feature requires the transmitter radio link option to be installed in both the vLoc3 series receiver and the Loc3 series transmitter.
Transmitter Control
(if the Transmitter Link option was ordered) (not available for the Loc-150Tx Transmitter)
- This
option is shown if the Transmitter Link is installed and active. This option will control the functions of the transmitter.
DFT
-
D
iscrete
F
ourier
T
ransform is a tool to help choose a frequency to apply to the target conductor. The DFT feature will aid
the user with nearby interference that may affect the locate quality.
New Survey
- Entering the Survey sub-menu causes a new survey initiated in the SD Card Data Logger.
DM Graphs are cleared so that a new survey is identifiable from older measurements.
Different Survey types can be selected as shown.
Navigating into this menu and selecting a survey type always causes a new survey to be registered, regardless of whether it is
the same type as the previous log record.
Electric
Gas etc.
Communications
Potable Water
Reclaimed Water
Sewer
Temporary Survey
Proposed Excavation
Oil
3.5.4
Setup - Informational
About
- This section holds the data about the locator, such as software revision, calibration data, etc.
Regulatory Labels
- This section shows the required FCC ID and IC information.
Self-Test
- The vLoc3 series receivers have a self-test feature that confirms the equipment is fit for use and that the calibration
has not drifted from its expected settings.
3.6 Self-Test
The self-test feature must be run in an area free from interference such as overhead fluorescent lighting, large transformers, etc.
Also, check that any nearby transmitters or sondes are switched off.
Select “Self-Test” from the user menu and press the “Return” button. Keep the receiver stationary while the test is running. After
a short while the test will complete and the unit will report Passed or Failed.
Self Test: PASSED
110.3
110.3
64.1
110.3
110.3
64.1
Self Test: FAILED
109.6
109.9
64.3
109.7
109.9
64.5
Examples of the self-test results
If the receiver fails the self-test try it again in an area with less interference. If it continues to fail return the unit to Vivax-
Metrotech or one of its approved repair centers for investigation.
Note that the vLoc3-DM low-frequency sensor foot is not tested during the self-test. To test this item
use the unit over a known pipe and check measured values fall within the published specification.