36
Speaker installation detail
Follow local regulations regarding securing these ceiling devices. If none apply, add a
safety wire (ceiling hanger wire) from the grid support to the device.
6. Install all speakers into selected locations.
All speakers have dual GRAY jacks, and there is no particular sequence for interconnection.
This allows for much flexibility in cable routing. Connect all of the devices in the procedure
room (LAB), including COMM speaker and the two MUSIC speakers (left and right) with
short white cables. The procedure room device nearest the console is connected to the console
at the
LAB WHITE
jack with a long white cable. Spare cables and connectors are provided. In
the control room, connect COMM and the two MUSIC speakers (left and right) with short blue
cables. Connect the control room speaker nearest the console to jack on the console labeled
TEK BLUE
, with a blue cable.
Using the provided color coded cables as instructed (white for procedure room and blue for
control room) will aid in the event that Vis-A-Vis customer support needs to be contacted to
assist with troubleshooting.
NOTE: Older devices having BLACK jacks cannot be used with this system.
Speaker
Support
Plate
Speaker
(Mounting Dogs
Extended)
Tile shown cut
away for clarity