33
6 INSTALLATION
6.1
Planing the Vitalinq™ installation
The Vitalinq™ has been designed for ease of installation. Please feel free to contact Vis-A-Vis
customer support at 800-319-6014 at anytime to assist with an installation.
There are microphones and speakers for the procedure room and for the control room. The
control room utilizes a desk microphone or a headset with foot switch (both are provided). The
procedure room microphone is what we refer to as the monitor microphone and is attached to
the face of the monitor or one of the monitors in the array opposite the physician
1
. All cables
are pre-terminated and may be run free or in conduit, subject to local regulations.
Using the layout in “6.2 Installing Vitalinq™ speakers” and the instructions in 6.1.1 and
6.1.2, select locations for the speakers and monitor microphone. Always check for adequate
space in the ceiling above the locations selected for the speakers.
Follow any applicable local
regulations regarding securing the speakers.
6.1.1 Procedure room (LAB)
In the procedure room the monitor microphone should be located on the video monitor directly
across from where the physician would typically stand and attached to the monitor with it’s
long axis oriented vertially. See “6.3.2 Monitor microphone placement”. For the physician to
hear best, the communication speaker should be near the physician and more than six feet away
from the microphone and the door to the control room. The music speakers should be placed
about four feet on either side of the physician, usually at the ends of the table.
Follow local
regulations regarding securing these devices.
6.1.2 Control room (TEK)
In the control room, the operator’s desk microphone should be placed in a position convenient
for the operator. For the best operation, the communication speaker should be more than six
feet away from the microphone and the door to the procedure room. The music speakers should be
placed in the ceiling, three or four feet on either side of the operator. If provided, the optional
auxiliary wireless headset should be located near the console.
Follow local regulations
regarding securing these devices.
6.1.3 Control room console
The cables from the speakers and monitor microphone need to connect to the rear of the desk
console. See layout typical layout on following page. If you are installing before the walls
are sheet-rocked, use a junction box near the console and one in the ceiling. Connect them
with a 1” conduit. Cables can be pulled through by staggering the connectors.
Follow local
regulations regarding securing these devices.
The communication system has volume controls for headset and overhead devices. Headset
jacks are on the front of the console and foot switch / desk microphone jacks are on the rear.
* If a special microphone referred to as the pyramid microphone was provided with the system you are
installing, please refer to the supplemental instructions which came with it for it’s installation. This
microphone is the size of and replaces a 2’ x 2’ ceiling tile.