VR9500 Hardware Manual
8
Connection
CONTINUED
Connecting the Device to a Network
NOTE:
Do not plug a phone jack connector into an RJ-45 port. This will
damage the Ethernet device. Use only twisted-pair cables with RJ-45 connectors
that conform to FCC standards.
·
Attach one end of a twisted-pair cable to the device's RJ-45 connector and the
other end to a switch or workstation computer.
·
Make sure the twisted pair cable does not exceed 328ft (100m) in length.
After connection, the network LED will turn green to indicate that the connection
is valid.
RJ-45 CONNECTOR
CAUTION
: The terminals on the unit should only be connected to passive
load transducers or source transducers that apply no more than 10V to the
circuit that originates from the VR9500 (or a value that equates to no more than
33Vac RMS/46.7V peak or 70Vdc total on the circuit).
CAUTION
: Other than the IEC inlet for input power to the unit, the terminals
on the rear of the unit should only be connected to SELV/double insulated
circuits of other equipment.
The 10/100/1000 BASE-T ports are copper
twisted-pair ports.
RJ-45 Pin Number Allocation for Ethernet Port
Pin No
Function
1
TxRx 1+
2
TxRx 1-
3
TxRx 2+
4
TxRx 2-
5
TxRx 3+
6
TxRx 3-
7
TxRx 4+
8
TxRx 4-
Содержание VR9500
Страница 1: ......
Страница 4: ......
Страница 5: ...1 support vibrationresearch com Hardware Overview Front and Back Panel Inputs Outputs ...
Страница 31: ......