19
1.12 A form pre-filled with the DMM details will appear. Fill in the remaining
fields with the appropriate values. Select
OK
.If the procedure fails to meet
the 1-year factory specifications, a warning message will appear. Check
your cable connections and procedure to confirm the proper steps were
followed. If the proper procedure was followed and the result is out of the
specifications, then the VR9500 must be recalibrated. Contact the factory
for recalibration information.
1.13 If the procedure meets the 1-year factory specifications:
a. Save data as
SN-TimeStamp-PreVerify
.
b. In the Microsoft Word document, add
As received
to the first line.
c. Run the full verification.