Electromagnetic Immunity
Immunity Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic
Environment - Guidance
Electrostatic
Discharge (ESD)
IEC 61000-4-2
±6 kV Contact
±8 kV Air
±6 kV Contact
±8 kV Air
Floors should be wood, concrete,
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%.
Electrical Fast
Transient/Burst
IEC 61000-4-4
±2 kV for power
supply lines
±1kV for
input/output lines
±2kV for power
supply lines
±1kV for
input/output lines
Main power quality should be that
of a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
±1 kV differential
mode
±2 kV common
mode
±1 kV differential
mode
Main power quality should be that
of a typical commercial or hospital
environment.
Voltage Dips, Short
Interruptions and
Voltage Variations
On Power Supply
Input Lines
IEC 61000-4-11
<5%
U
T
(>95% dip in
U
T
)
for 0.5cycle
40%
U
T
(60% dip in
U
T
)
for 5 cycle
70%
U
T
(30% dip in
U
T
)
for 25 cycle
<5%
U
T
(<95% dip in
U
T
)
for 5s
<5%
U
T
(>95% dip in
U
T
)
for 0.5cycle
40%
U
T
(60% dip in
U
T
)
for 5 cycle
70%
U
T
(30% dip in
U
T
)
for 25 cycle
<5%
U
T
(<95% dip in
U
T
)
for 5s
Main power quality should be that
of a typical commercial or hospital
environment. If the user of the
V1407 requires continued
operation during power main
interruptions, it is recommended
that the V1407 be powered from
an uninterruptible power supply or
a battery.
Power Frequency
(50/60 Hz)
Magnetic Field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.
NOTE:
U
T
is the A.C. main voltage prior to application of the test level.