![Vetland V1407 Скачать руководство пользователя страница 51](http://html.mh-extra.com/html/vetland/v1407/v1407_user-manual_1010096051.webp)
V1407 User Manual
51
Arrhythmia Guide
The arrhythmia guide is not species specific and is only a tool for comparison.
Name of Arrhythmia
Description
Tachycardia (TAC)
Heart rate is over 140 bpm for a minute.
Bradycardia (BRD)
Heart rate is under 40 bpm for a minute.
Premature Ventricular
Contraction (PVC)
Abnormal heart rate is over 1.
Ventricular Tachycardia (VTAC) Over 8 PVC is over 140 bpm
Asystole (ASY)
No electric action of heart.
Bigeminy (BGM)
After normal heart rate once, two times or more PVC
Trigeminy (TGM)
After normal heart rate at two times, two times or more PVC.
Ventricular (VENT)
Check the state of the ventricular
Ventricular Fibrillation (VFIB)
A fibrillatory wave for 4 consecutive seconds
Couplet (CPT)
Abnormal heart rate is two times
Triplet (TPT)
Abnormal heart rate is more than three times
Missed Beat (MIB)
ECG wave is missed intermediately
R-on-T
PCV is detected before finishing T wave in the normal ECG
wave.
NOTE
The V1407 provides 13 types of arrhythmia display in two alarm levels:
- High Alarm: VTAC, ASY
- Low Alarm: PVC, BGM, TGM, TAC, BRD, VENT, VFIB, CPT, TPT, MIB, R-on-T
If you press the
button during the arrhythmia alarm, the
icon is displayed on the
top of the screen and all alarms will be temporarily silenced for 2 minutes.
Although studied widely in human medicine, arrhythmia measurements have not been
validated in veterinary medicine and are offered only as an ancillary tool.