Natural Lifetime for Spontaneous Battery Discharge
The table below shows the lifetime for spontaneous discharge of the battery when the product
is unused.
Storage Temperature
(
◦
C)
Life for Spontaneous Discharge
(Months)
0– 20
9
21– 30
6
31– 45
3
Type of Battery
- Spec: 11.1 V, 2200 mA
- Type: Lithium Ion Battery
Time for Battery Exchange
The lifetime of the battery gets shorter if it is not 100% recharged for use.
The charge / discharge ration is described in the table below.
Charge / Discharge Ration
Number of Times Used
100 %
1200
50 %
500
30 %
200
The battery is recharged when AC power is confirmed and the main switch is on at the back of
the monitor.
How to Change the Battery
The life expectancy of the battery is nearing its end when 30% charge /discharge ration is
occurring. Please contact Vetland Medical to arrange delivery of a replacement battery pack.
WARNING
The battery is easily exchanged via the battery access port. Do not disassemble the monitor as
damage may occur.