ECG Lead Change
1
To change the ECG1 lead waveform setting, press
Menu button one time on the Main screen.
Select ECG
LEAD
CH 1
and then select the
lead.
2
Change the ECG2 and ECG3 lead waveform settings using the same method used in Step 1,
except select CH 2 and then CH3
.
3
By operating as detailed above, you can select each function of the ECG mode.
Setting
Mode
Range
Description
LEAD
LEAD I, II, III, aVR, aVL, aVF, V
Sets the type of ECG lead.
GAIN
2.5, 5, 10, 20, 30, 40 mm/mV, AUTO
Sets the size of the ECG wave.
SPEED
6.25, 12.5, 25, 50 mm/sec
Sets the velocity of the ECG wave.
CONFIG
CABLE
3 LEAD, 5 LEAD
Sets the type of ECG cable.
NOTCH
START, STOP
Synchronizes the monitor with the frequency of
the power being supplied.
AREA
50Hz, 60 Hz
Establishes the frequency being used by the
monitor’s power input.
FILTER
0.5 – 40 Hz
Sets the analog and digital filter for the ECG
wave.
0.5 – 80 Hz
0.05 – 40 Hz
0.05 – 80 Hz
MONITORING
ST
J + 0, 20, 30, 40,
50, 60, 70, 80 ms
Sets the J point.
ARRHYTHMIA
START, STOP
Sets the usage of arrhythmia analysis.
PACEMAKER
START, STOP
Sets the detection of the pacemaker.
HR CALC.
4 – 16
Sets parameters to compare for the HR Calc.
DEFIB. SYNC
ON / OFF
Sets synchronization when using ECG OUT.
QRS WIDTH
100,200,300,400,
600, 800 ms, AUTO
Sets the usage of HR detect.