VENOUS STAT Function
This function is used to easily find a patient’s vein by controlling the flow of the vein by maintaining
constant pressure to the cuff.
1
Press the Menu button once and select, NIBP
PRESSURE
VENOUS STAT and set the pressure
value.
2
To exit the VENOUS STAT mode, press NIBP.
Caution
The VENOUS STAT function is only for patients whose veins are difficult to find. Avoid
using this function for normal patients.
After using the VENOUS STAT function, set to AUTO or MANUAL for measurement.
Note
VENOUS STAT controls flow of the vein for 170 seconds for VET_ Large, and 85
seconds for VET_Small.
VENOUS STAT is able to be set to between 50 – 200 mmHg (5mmHg unit). The default
setting is 100 mmHg.