V1407 User Manual
87
- DC OFFSET: -300mV to 300mV
- ARRHYHMIA display (TAC, BRD, PVC, VTAC, ASY, BGM, TGM, VENT, VFIB, CPT, TPT, MIB,
R on T)
- ST Analysis: Range: -9.9mm – +9.9mm, Resolution: 0.1mm
- Pacemaker Detect
- HR Calculation: 4 – 16 waves
- Defibrillator Sync
- QRS WIDTH: 100, 200, 300, 400, 600, 800ms, AUTO.
Heart Rate
- Measurement Range: 0 – 300bpm
- Accuracy: ± 2bpm (Measurement value)
SpO2 (Pulse Oximetery)
- Measurement Range: 0 – 100%
- Accuracy: 70 – 100% SpO2: ± 2%, 50 – 69% SpO2: ± 3%
0 – 49% SpO2: Unspecified
- Alarm Range:
High limit - 0 – 100%, OFF, Low limit - OFF, 0 – 100%
- Pulse Range: 0 – 300bpm
- Pulse Measurement Accuracy: 20 – 240bpm ±2 bpm
241 – 300bpm ±3 bpm
NIBP (Non-invasive Blood Pressure Measurement)
- Measurement Method: Oscillometric, Automatic measurement
- Display Value: Systolic, Diastolic, Mean Pressure
- Measurement Range: 0 – 300 mmHg
- Accuracy: ±8 mmHg (SYSTOLIC, DIASTOLIC), ±5 mmHg (MEAN)
- Alarm Range:
NIBP
[mmHg]
Display
High
Low
Systolic
0 – 275, OFF
OFF, 0 – 275
Diastolic
Mean
- Auto Pressure: Manual, Automatic (1, 2, 3, 4, 5, 10, 15, 20, 30 min. and 1, 2, 4, 8, 12 hour intervals)
- NIBP STAT: 5 – 15 min
- NIBP Max. Limit: 70 – 300mmHg