Ds500, Ds500+ Operating manual
Original instructions (en)
12.2 Error Codes
On screen error codes provide useful information in the event of an error:
Error ID
Message
Solution
10002
Pump Head Issue
•
Make sure cartridge is properly and securely fixed.
•
Replace the cartridge if burst has been identified.
11500
Internal Memory Error
•
Power cycle the unit
11501
External Memory Error
•
Power cycle the unit
12008
Motor Failure
•
Power cycle the unit - if problem persists, contact Verder
for technical support.
12016
Motor Failure
•
Power cycle the unit.
•
Remove the cartridge (see 10.3.3 Cartridge information)
and confirm if the pump resumes normal operation.
If yes - install the same cartridge to see if normal operation
resumes. If no - fit a new cartridge and re-check if normal
operation resumes.
•
If unit has been running for a prolonged period, stop the
motor before re-energising it.
12702
Fluid Low
•
Check fluid level or re-adjust the warning% - see 10.3.4
Table 17. - Error codes